The «Digital Inputs» display shows the assignment of
digital inputs 1 to 4. Press
to get to the
«Digital Inputs 5 to 8» display.
to return to the «Configuration» display.
The «Digital Inputs» display shows the assignment of
digital inputs 5 to 8. Press.
to get to the
«Digital Inputs 1 to 4» display.
to return to the «Inputs/Outputs» display.
5.2.11
Digital Outputs
The CMOVE 1250 has eight digital outputs. This menu informs on the assignment and
status of the digital outputs but it does not allow making any settings, except for DO 3. If
there is a black square at the right end of a line, the corresponding signal is active
(Operation via inputs and outputs see chapter «5.4 Operation Via Inputs and Outputs»).
The «Configuration» display shows the current
settings of the operating mode, valve type, pressure
sensor, and nominal value (setpoint) with source
(I = internal, E = external). Press
to get to the «Inputs/Outputs» display.
DI5-8
DI5-8
DI 1: Flow decrease
DI 2: Flow increase
DI 3: External CLOSE
DI 4: External OPEN
Digital Inputs
DI1-4
DI1-4
DI 5: Flow mode
DI 6: Pressure mode
DI 7: Emission ON
DI 8: Degas ON
Digital Inputs
I/O
VALVE SENSOR I/O GEN.
Mode :
Pressure
Valve :
MOVE1250STD
Sensor : CTR1000T
Nom. I : 1.33E+01 mbar
Configuration