DO 3 Deviation
The settings of this parameter are only effective in the Pressure control
mode.
«DO 3 Deviation» is a control output. Its settings do not directly affect the pressure control
process or the gas flow adjustment. They only define the conditions under which a
superordinated control system considers the valve to be "in position".
Figure 5-6
This parameter is set automatically to 10% of the F.S. (full scale) when an
implemented pressure sensor (see
[3] … [12]) is connected and
selected in the configuration menu.
The «Digital Outputs» display shows the assignment
of digital outputs 1 to 4. Press
to get to the
«Digital Outputs 5 to 8» display.
to get to the
«DO 3 Deviation» display.
to return to the
«Inputs/Outputs» display.
Signal OFF [% FS]
DO signal with
ON-delay t
DO signal without
ON-delay
Nominal value (setpoint)
Signal ON [% FS]
Signal ON [% FS]
Signal OFF [% FS]
Pressure
X
OFF
OFF
DO 3
DO 3
ON
ON
t
X
t
X
t
X
Time
Time
Time
DO5-8
DO3
3
O
D
8
-
5
O
D
DO 1: Valve close
DO 2: Valve open
DO 3: in position
DO 4: Valve error
Digital Outputs