4.2.4
Pressure Sensor
NOTICE
Malfunctions
Connecting more than one pressure sensor at a time may cause malfunctions.
Connect only one sensor at a time.
The CMOVE 1250 is equipped
with three different sensor
receptacles (compatible pressure
sensors see chapter «12»).
Figure 4-5
Pin assignment X5 D-Sub receptacle
X5 sensor
Designation
12
Signal
Pin
Emission status input
+24 VDC
1
Signal input (measurement signal)
0 … +10 VDC
2
Warning status input
+24 VDC
3
Control output Emission ON
13
+24 VDC
4
Supply common
GND
5
Not connected
6
Degas ON control output
+24 VDC
7
Supply
+24 VDC
8
Not connected
9
Not connected
10
Not connected
11
Signal common
GND
12
Not connected
13
Not connected
14
Shielding of housing
GND
15
Table 4-3
12
The pin assignment depends on the pressure sensor used. See connection instructions of the
corresponding pressure sensor (→
[3] … [12]).
13
Pressure sensors specified for a max. operating pressure (e.g. ITR 100) will start operation only after
the signal "Emission ON" is applied to the digital input DI 7 of the CMOVE 1250 (see chapter
«5.4 Operation Via Inputs and Outputs»).
X9 valve
X4 sensor
X5 sensor
X2 sensor
X5: D-Sub15-pin
X2: DIN 6-pin
X4: FCC68 8-pin