602494EA.DOCX
17200234_002_C1/2016 - © Leybold
23/102
4.2.2
Mains Power Connection
DANGER
Electric shock
Incorrectly grounded products can be extremely hazardous in the event of a
fault.
Use only a 3-conductor power cable with protective ground. The power
connector may only be plugged into a socket with a protective ground. The
protection must not be nullified by an extension cable without protective
ground.
NOTICE
Wrong voltage
Electrical components may get damaged.
Make sure that the mains voltage is in conformity with the voltage specified on
the controller; see chapter «1.4 Technical Data».
The unit is supplied with a 2 m power cord.
If the mains cable is not compatible with your system, use your own, suitable cable with
protective ground.
The socket must befuse-protected with
10 A
max
Figure 4-2
The mains voltage should be supplied and turned on via a central power
distributor because the main switch of the CMOVE 1250 cannot be reached
from the front.