Diagnostics and troubleshooting
Leuze electronic
BPS 348i
87
Table 10.6:
Position measurement errors – causes and measures
Error
Possible cause
Measures
Measurement value
or read quality is
continuously instable
• Soiling of the BPS optics
• Clean the optics of the BPS
Measurement value
or read quality is
poor
• at certain posi
-
tion values
• always at the
same position
values
• Soiling of the bar code tape
• Clean the bar code tape
• Replace the bar code tape
No measurement
value can be deter
-
mined
• No code in scanning beam
• Code not in the working range
of the BPS
• Align the scanning beam with the
bar code tape
• Align the BPS with the bar code tape
(working range 50 mm - 170 mm)
Faulty measure
-
ment value
• Wrong bar code tape
BCB grid different from BPS
configuration
• Change BPS configuration to the
bar code tape that is being used