The fluorescence attachment contains a filter-slide

housing for two suppression filters. The standard

equipment contains one suppression filter. The

second position can be used for an additional

suppression filter.

Filter-slide housing with filter glass and clear

glass. (The second filter shown is optional).

If only one suppression filter is used, a clear glass

filter must be inserted in the second position. This

glass is required so that the working distance

remains unchanged if the user switches between

fluorescence and bright field application.

The filter-slide housing is mounted on the 

stereomicroscope as shown in Fig. 2.

The left side features a threaded hole with 

recessed screw that must be tightened 

using an Allen key (1.5 mm).

The left holder with the red UV protective filter is

optional. Leica recommends the use of such a filter

to protect the eyes of the observer (Fig. 3).

Hinged arm see page 12.

Fig. 1 - Suppression filter with 

filter-slide housing

Fig. 2 - Overview of the 

assembly of the S series

Fig. 3 - Connection of the 

hinged arm on the microscope

carrier S series

Assembly

Summary of Contents for L5 FL

Page 1: ...Leica L5 FL User manual...

Page 2: ...omicroscopes M series 13 Stereomicroscopes S series 14 Remote control connection 15 Lamp Installation replacement 16 Starting up 17 Luminous intensity setting 18 Electronic setting 18 Mechanical diaph...

Page 3: ...handle 10 Lamp housing with halogen cold light reflector lamp 11 Release button for lamp housing 12 Power supply cord 13 Ventilation openings front side of instrument 14 Ventilation openings side of...

Page 4: ...ooms only see page 25 Technical Data This instrument is not intended for operation in explosion prone areas A safe disconnect from the power supply system is achieved by disconnecting the power plug T...

Page 5: ...teristic of converting incident light into heat damages may occur to heat sensitive or inflammable light absorbing materials To avoid such thermal damages and possible fire hazards or risk of burns pl...

Page 6: ...e illumination to the absolute necessary measure It is absolutely necessary that the Leica L5 FL cold light source is operated with the voltage listed on the identification label 17 all ventilation op...

Page 7: ...el at risk Caution observe the documentation Nonobservance can cause malfunctioning or damage to the equipment Warning about hot surfaces Risk of burns in case of a hot surface Important Information T...

Page 8: ...8 Leica Stereomicroscope MZ6 with Leica L5 FL Leica Stereomicroscope S8 APO with Leica L5 FL...

Page 9: ...itions and remove it The adapter is inserted in reverse order Please store the connecting adapter in a safe place for later use with other fiber optic light guides Screw the black ring back on the fib...

Page 10: ...atures position marks the position 0 through 4 are marked by means of raised points in corresponding numbers The excitation filter is inserted in the Leica L5 FL cold light source To insert the filter...

Page 11: ...The daylight filter is part of the standard delivery and is used to increase the color temperature for bright field applications The filter wheel features 5 positions for max five filters Fit the fil...

Page 12: ...freely movable The support arm can now be positioned Next turn the rotary knob clockwise to lock all joints Hinged arm for illumination and UV protective filter Stereomicroscopes M series Connection o...

Page 13: ...suppression filters The standard equipment contains one suppression filter The second position can be used for an additional suppression filter Filter slide housing with filter slide and filters One...

Page 14: ...required so that the working distance remains unchanged if the user switches between fluorescence and bright field application The filter slide housing is mounted on the stereomicroscope as shown in F...

Page 15: ...ntensity control 3 is deactivated Remote control and tilt switch allow switching the light source to STANDBY mode to protect the halogen lamp A corresponding note 21 appears on the display Please alwa...

Page 16: ...housing 10 by pressing the release button 11 and pull it out to the stop Fig 1 Press both levers of the special socket down and pull out the defective bulb The two levers must also be pressed down whe...

Page 17: ...ec ted fiber optic light guides Position The instrument is switched off Position The instrument is switched on The Leica L5 FL is equipped with a soft start to protect the halogen lamp that reduces an...

Page 18: ...so results in a change of the color temperature of the emitted light The mechanical adjustment allows for the continuous change of the luminous intensity with constant color temperature Four lock in p...

Page 19: ...lb manufacturer type dependent The LCD display 5 shows the approximate color temperature of the light emitted by the halogen lamp 18 The color temperature of the light can be adjusted by turning the l...

Page 20: ...y opened diaphragm achieves the maximum brightness Turning the adjusting knob from one lock in position to the next results in approximately 2x increase or 1 2x decrease of the luminous intensity The...

Page 21: ...en dent upon the magnification and the working distance of the stereomicroscope The focusing lens features a slide mechanism to obtain a perfect adjustment of the light spot diameter to the field of v...

Page 22: ...Use a soft dry cloth or commercial plastic clean sing wipes to clean the outside of the instrument The display features an error status display 19 Malfunctions can be detected quickly and easily Err...

Page 23: ...ment with switched off light source Lamp off fan operating Bulb defective Replace bulb see p 16 error status display No bulb Insert with light source Err1 switched off Lamp off fan operating Insuffici...

Page 24: ...bulb type that will allow for perfect light utilization and illuminated area is deliver ed The following functions can be controlled via remote control Light on off STANDBY Electronic luminous intens...

Page 25: ...e C 5 40 Relative humidity up to 31 C ambient temperature 85 from 31 C to 40 C ambient temperature linear decrease to 75 Atmospheric pressure hPa 700 1060 Transport and storage Temperature C 40 70 Rel...

Page 26: ...1 71 726 33 33 Fax 41 71 726 33 34 www leica microsystems com www stereomicroscopy com Illustrations descriptions and technical data not binding Subject to change without notice Leica Microsystems AG...

Reviews: