When NIBP is selected in the main screen, NIBP Setup window will pop-up on the
screen as shown in Figure 4.8.
NIBP
Mode
:
Select this to choose Manual, Stat or Auto mode from
drop down list.
Manual
:
For taking individual BP readings. After selecting this function,
press start NIBP measurement key on front panel to begin manual
NIBP measurement. To abort NIBP measurement midway, press
the same key again.
Auto
:
For taking BP readings automatically at set intervals (Timer
set). Press START on the front panel to take the first reading.
Next NIBP reading starts after the set time interval. The time
left before the next measurement to begin is displayed on screen.
This displayed time is decremented every minute. Timer can
be set only if Automatic is selected.
Stat
:
Allows the user to take as many readings of NIBP as possible in
5 minutes of continuous operation.
Alarm
:
Select this to enable (switch ON) or disable (switch OFF) the
NIBP alarm.
Sys High Limit
:
Select this to set upper alarm limit for Systolic alarm from Numeric
keypad ( To use this function, ensure that alarm is enabled).
Figure 4.8
4.7
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...