CAPNOGRAPHY
Theory of operation
The principle of Capnography is based on the absorption of infrared radiation
by CO
2
. The technique is known as non-dispersive infrared absorption technique.
The spectral region is particularly appropriate for measuring carbon dioxide
because it has a strong absorption band in the near Infrared wavelength. CO
2
selectively absorbs specific wavelengths of Infrared light. The main part of the
system is CO
2
bench, which consist of one IR source and detector. A sample
of patient's expired gas that consists of CO
2
, Water vapors, Nitrogen, Anaesthetic
Agents, etc. is drawn from a lightweight T-piece through a sample line into the
CO
2
bench. Expired air is aspirated into measuring chamber (CO
2
bench) by a
small pump. The amount of light passing through a sample cell varies according
to the concentration of CO
2
in the sample cell. When concentration of CO
2
in
sample cell is high, more light is absorbed by the sample and therefore a small
amount of light reaches the detector as compared to low concentration of CO
2
.
The amount of light absorbed is proportional to the concentration of CO
2
. The
CO
2
concentration measured by the monitor is usually expressed as end tidal
concentration of CO
2
(EtCO
2
), expressed in terms of mmHg or percentage (%)
or Kilopascal (Kpa).
Capnogram - Nature of waveform
CO
2
waveform (Refer Figure 2.16) reflects various stages in breathing. Capnograph
is an important diagnostic tool because its shape is virtually identical in all
basically healthy people.
Figure 2.16
2.16
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...