Menu
Select this to go to Monitor settings / System functions.
Trend
Select this to retrieve the data of the patients recorded.
Display
Select this to choose different display formats.
Volume
Select this to set and vary alarm volume and beep volume.
Recall
Select this to recall alarm and ECG trend.
Autoset
Select this to set alarm limits automatically for all the
parameters except NIBP.
ECG :
HR/PR high limit
=
(Present HR/PR) x 8/10 +76
HR/PR low limit
=
(Present HR/PR) x 8/10 +2
SpO
2
:
SpO
2
high limit
=
100
SpO
2
low limit
=
(Present SpO
2
Value -8) or 80
whichever is greater.
IBP :
Systolic high limit
=
Present Systolic value +8
Systolic low limit
=
Present Systolic value -8
Diastolic high limit
=
Present Diastolic value +8
Diastolic low limit
=
Present Diastolic value -8
Respiration :
Respiration high limit = Present Respiration rate +5
Respiration low limit = Present Respiration rate -5
Temperature :
T1 high limit
=
Present T1 value +2
T1 low limit
=
Present T1 value –2
Capnography :
EtCO
2
high limit =
Present value +5
EtCO
2
low limit
=
Present value -5
FiCO
2
high limit
=
Present value +2
FiCO
2
low limit
=
0
3.6
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...