Note
Do not rely on the displayed HR value when pacer
is OFF and pacer pulses are detected as QRS.
Beep source
:
User selectable beep source from ECG/IBP/
SpO
2
/NIBP
Leads
:
3 leads for 3 lead cable.
5 leads for 5 lead cable
QRS Indicator
:
Beep and flashing heart symbol for every
QRS complex detected.
9. Heart Rate
Range
:
20-350 BPM
Accuracy
:
2BPM or 2% which ever is greater
HR alarms
: Adjustable alarm limits
Min : 20 to 320 BPM.
Max : 50 to 350 BPM
4 seconds delay for HR alarms
Time between the silencing of the alarm
and activation is 14 seconds.
Heart rate meter response time: Change from 80 to 120 BPM (6 to 7 sec).
Change from 80 to 40 BPM (11 to 12 sec).
Pacemaker pulse rejection/
:
When pacer is ON the monitor will display
detection capability
the HR for all single and double pacemaker
pulses either 150 or 250 msec apart with a
pulse width of 0.1 to 2msec and amplitudes
±2 mV to ±500 mV without overshoot.
Lower threshold of slew rate for detection of
pacer pulse is 2.2 V/s.
10. Respiration: (When source is from ECG)
Input
:
From ECG cable ( RA - LA / RA - LL)
Leakage current
:
<10
µ
A at 240V AC, 50Hz
Excitation current
:
< 300
µ
A at 50 kHz.
Sensitivity (max)
:
0.2 Ohms/cm
Sensitivity (min)
:
4 Ohms/cm.
Range
:
0 -150 breaths\min.
9.6
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...