Auto zero
:
Zero pressure reference is automatically
established after every reading
Cuff inflation
Initial inflation
Adult
:
160 mmHg
Neonates
:
90 mmHg
Pediatric
:
120 mmHg
Subsequent inflation approximately 30 mmHg greater than previous systolic pressure
Cuff deflation
:
Automatic
Alarms
:
Adjustable alarm limits (for both Sys and Dia)
Upper
:
30 to 250 mmHg
Lower
:
20 to 240 mmHg
Safety features
:
Automatic deflation if, cuff pressure exceed.
•
300 mmHg (Adult and Pediatric mode)
•
150 mmHg (Neonate mode)
•
Measurement time exceeds more than one minute to deflate.
15. Drug calculation
25 Standard default drug details are as follows:
Aminophylline, Diltiazem, Dobutamine, Dopamine, Epinephrine, Esmolol HCL,
Fentanyl Citrate, Heparin Sodium, Inamrinone Lactate, Insulin Regular, Isoproterenol
HCL,LabetolHCL,Lorazepam,Midazolam,Milrinone,Nicardipine,Nitroglycerin,Nitroprusside,
Norepinephrine, Phenylephrine,Procainamide HCL, Propofol, Vasopressin,
Lidocaine HCL and Pitocin.
Apart from above drugs there is a provision to add 10 New drugs.
16. Pressure IBP1-IBP4
Input
:
Isolated and floating. Protected against surges
produced by Electro-surgical and
defibrillator potential with L&T approved
cables.
IBP1-IBP4 w/f sampling rate
:
100
Input impedance
:
>1M
Ω
Leakage Current
:
<10
µΑ
at 240V AC, 50Hz
Processed Pressure Signals
:
Systolic, Diastolic and Mean.
9.10
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...