Capnography Maintenance Schedule
•
Use specified accessories only.
•
Check the accessories before use.
•
Sample line is disposable and for single use only.
•
Incase of any water condensation in sample line, replace the sample
line with a new one.
•
CO
2
module should be calibrated once in a year or 4000 hours of usage, which
ever is earlier. This calibration should be done by authorized L&T personnel.
The calibration must be performed with a manufacturer approved calibration
kit. The manufacturer approved calibration kit must be purchased from Scott
Medical. Calibration gas contains 5% CO2, 21% O2 and balance N2. Switch
ON the Monitor.
Procedure for calibrating Capnography Module
•
Switch ON the monitor.
•
Go to set up screen by pressing optical encoder during self test.
•
Press
CO
2
CAL
from the
Setting
menu.
•
Connect filter line from calibration kit to the monitor.
•
Press start
CAL
and
Confirm
for further proceedings.
•
Message " Keep pressing cylinder knob to deliver the gas" will display. Press
the knob of the cylinder kit and select
OK
.
•
Message "Calibrating... Please wait" will display.
•
Message "Calibrating ... Please remove gas" will display.
•
After calibration , message "Calibration succeded" will display
•
Remove filter line from the monitor.
Check the last calibrated date in Menu System All Info
6.5
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...