Leak Tightness
The leak rate of the module flow system is less than 40 mBr per minute
when a 30% vacuum is invoked on the flow system.
FilterLine Recognition Safeguard (FRS)
The "FRS" enables the module to detect the presence of the FilterLine at
startup and in normal operation. The pump doesn't draw in gas from the input
port when a FilterLine is not connected. While in normal operating mode the
"FRS Lock" detection time is no longer than 1 second.
Purge
The module performs a purging whenever it detects an occlusion in the
FilterLine or in the airway adapter. The module informs the host of the purge
situation by setting the "Purging in progress" bit in the Wave Message While
purging, the CO
2
values are invalid. The duration of the purging is up to 30
s e c o n d s .
The module recovers from occlusion when FilterLine is replaced. The module
then shifts to initialization mode for typically less than 5 seconds and then goes
to normal measurement mode.
Definitions
Breath
:
A r i s e a n d f a l l i n t h e c a r b o n -
dioxide concentration of at least
one percent carbon-dioxide is no
less than 0.4 seconds.
One inha One exhalation
= One Breath.
Respiration Rate (RR)
:
Number of Breaths per Min.
2.20
Summary of Contents for Star 55
Page 1: ......
Page 18: ...Star 55 1 1 DISPLAY CONTROLS CONNECTIONS AND OUTPUTS ...
Page 19: ...Figure 1 1 23 15 19 16 17 22 18 21 20 1 234 5 6 7 8 9 10 11 12 13 14 Front Panel 1 1 ...
Page 22: ...Star 55 Left blank intentionally 1 4 ...
Page 23: ...Star 55 0843 1 2 Figure 1 2 Rear Panel 1 5 3 4 ...
Page 25: ...Left Side Panel Figure 1 3 4 7 6 8 9 3 2 5 1 1 7 ...
Page 27: ...Figure 1 5 1 Right Side Panel 2 3 1 9 ...
Page 36: ...Star 55 2 APPLICATION GUIDE 2 ...
Page 51: ...Star 55 Figure 2 15 2 15 ...
Page 59: ...Star 55 OPERATIONS AND MAIN SCREEN 3 3 ...
Page 61: ...3 2 MAIN SCREEN 1 Figure 3 1 Note Main screen when maximum i e eight waveforms are present ...
Page 62: ...Star 55 MAIN SCREEN 2 Figure 3 2 Note Main screen when maximum ECG waveforms are present 3 3 ...
Page 73: ...Star 55 4 PARAMETER SETTING 4 ...
Page 91: ...Star 55 SYSTEM FUNCTIONS 5 5 ...
Page 103: ...Figure 5 13 5 12 ...
Page 114: ...Star 55 6 MAINTENANCE SCHEDULE 6 ...
Page 121: ...Star 55 ACCESSORIES 7 7 ...
Page 130: ...Star 55 8 TROUBLE SHOOTING AND WAVEFORM 8 ...
Page 150: ...Star 55 TECHNICAL SPECIFICATIONS 9 9 ...
Page 166: ...Star 55 10 WARRANTY AND WARRANTY SERVICE 10 ...