3.5
Model Type Definitions
BG3K1
3.5.1
The BG3K1 measures CH
4
, CO
2
, and O
2
as standard with a choice of up
to two additional gases (one internal to the module and one external)
and measures from one sample point.
It has user selectable sample and air purge intervals that are exhausted
back to the process and atmosphere respectively.
The 4-20mA and Modbus outputs are updated at the end of each
sample.
The following image is a simple block diagram of the system including
the optional external gas sensor:
BG3K2
3.5.2
The BG3K2 measures CH
4
, CO
2
, and O
2
as standard with a choice of up
to two additional gases (one internal to the module and one external)
and measures from two sample points.
It has user selectable sample and air purge intervals that are exhausted
back to the process and atmosphere respectively.
The 4-20mA and Modbus outputs are updated at the end of each
sample.
The following image is a simple block diagram of the system including
the optional external gas sensor: