Figure 10 - 4-20mA notification
Indicates current sample point solenoid – This mode enables a relay for
when a sample point is being monitored. This is useful if the operator
needs to know when a sample point is being monitored
Figure 11 - Sample point notification
Note: When enabling the relays for 4-20mA or sample point notification,
the remaining available relays will automatically be defaulted to ‘Alarms’.
4)
Press key ‘2’ to define whether a relay is used to indicate when an air
purge is occurring. If this option is selected, this will automatically be
defaulted to ‘Relay 5’.
5)
Press key ‘3’ to define whether a relay is used to indicate when the
catchpot contains liquid and needs emptying. If this option is selected, this
will automatically be defaulted to ‘Relay 7’.