114
Operating Manual
General Operating Instructions
OMBG3000 – Iss.01
© Copyright 2016
Screen 29 - Restore to Factory settings
2)
Press the right soft key to ‘Abort’ which does not reset the calibration data
and will return the operator to the calibration menu.
3)
Press the left soft key to ‘Continue’, which resets the calibration data and
will return the operator to the calibration menu.
Note: Restoring the system to factory settings will stop the sampling
process and freeze the outputs at their last known value.
5.7.3.3
Fit New Cell
This option should be selected when the user has installed a new pre-
calibrated external cell or external sensor module (cell complete with PCB). It
over-writes the current factory calibration values with the data provided with
the cell.
Note: Selecting this option will stop the sampling process and freeze the
outputs at their last known value.
1)
From the ‘Calibration’ menu press key ‘3’ to select the ‘Fit New Cell’
option.