background image

13

1.  From the Service menu, tap the 

Valve Calibration

 button.

Calibrating Brand Syrup Modules - Using a Graduated Cylinder

 

Ensure there is ice on the cold plate and the lines are 

cold before attempting to set the flow rates on the 

valves. The drink temperature should be no higher  

than 40°F (4.4°C) when flow rates are set. 

NOTE

2.  Tap the 

Calibrate

 on the Syrup Valve to be calibrated. 

3. 

Set the Timer to the ON position and select Volume as the 

desired unit of measurement.

4.  Using the slider, enter in an amount of 

50 ml 

as the 

preset dispensing amount. 

6.  Examine  the  dispensed  volume  in  the  graduated  cylinder.             

If the dispensed volume does not match the value of 50 ml, 

use a screwdriver to adjust the brand syrup flow control. 

(See Plumbing Diagram on page 26 for reference). 

The water flow rate was set to 73.93 ml/sec, which 

makes the final syrup flow rate 14.79 ml/s. The finished 

drink flow rate dispensed will be 88.72 ml/s. 

NOTE

A

B

C

D

Increase

Decrease

A. Flow Control

B. Valve Retainer 

C. Solenoid

D. Valve Body

7.  Repeat steps 5 and 6 until the designated volume of 50 ml is 

achieved.

8. 

Repeat steps 2 - 6 for the remaining brand syrup modules.

5.  With the graduated cylinder placed in a position below the 

nozzle, tap the 

Purge

 button. The unit will  

dispense the designated syrup amount. 

 

Assurez-vous qu’il y a de la glace sur la plaque froide 

et que les conduites sont froides avant de tenter de 

régler les débits sur les vannes. La température de la 

boisson ne doit pas dépasser 40 ° F (4,4 ° C) lorsque 

les débits sont réglés.

REMARQUE

 

Le débit d’eau a été réglé à 73,93 ml/s, ce qui fait que le 

débit de sirop final est de 14,79 ml/s. Le débit de bois

-

son fini distribué sera de 88,72 ml/s.

REMARQUE

Fill to 50 ml

Graduated Cylinder

Summary of Contents for Twin Pour 30

Page 1: ...not be used by children or infirm persons without supervision This appliance can be used by children aged from 8 years and above and persons with reduced physical sensory or mental capabilities or la...

Page 2: ...hez pas cet appareil dans une prise lectrique multiple partag e par d autres appareils Ne placez pas plusieurs prises d alimentation multiples portables ou alimentations lectriques portables l arri re...

Page 3: ...an air gap a backflow prevention device or another approved method to comply with NSF standards A leaking inlet water check valve will allow carbonated water to flow back through the pump when it is s...

Page 4: ...aining three sides Level unit if necessary For optimum perfor mance place the unit at a 0 tilt The maximum tilt is 5 Leveling the Dispenser NSF listed units must be sealed to the counter NOTE To ensur...

Page 5: ...at se trouve dans le bac gla ons Durant le cycle d agitation automatique et lors de la distribution des gla ons veiller ce que l espace soit suffisant entre le dessus de l amas de gla ons et le bas de...

Page 6: ...nect tubing to all syrup inlets See Plumbing Diagrams on page 26 for reference NOTE 12 Route drain hose from designated open type drain to fitting on Drip Tray and connect hose to fitting if applicabl...

Page 7: ...E 13 Attach fan bracket to fan assembly using screws provided Condensation Fan Installation 14 Install condensation fan on the underside of the unit by 1 pushing fan bracket into slots in the right wa...

Page 8: ...OT TURN ON CO2 SUPPLY AT THIS TIME WARNING 7 Repeat Step 6 for both low pressure CO2 regulators on the regulator manifold routed to the unit and the syrup pumps 5 Connect tubing routed from the tee at...

Page 9: ...the service menu for drink setup must be accessed 10 Access the Service Menu by turning the keyswitch clock wise to the third position 6 Turn on the power to the dispenser by pressing the on off toggl...

Page 10: ...oth of the low pressure regulators on the regulator manifold to 75 psi 0 517 MPa then tighten locknut with wrench 22 Activate each soda module until gas out 23 Plug in the remote carbonator pump deck...

Page 11: ...button on the lower left corner to return to the Valve Configuration Screen 7 Repeat Steps 3 and 5 for any of the other brand or flavor modules 8 Tap the Back button to return to the Service menu 12 T...

Page 12: ...bration button 2 Tap the Calibrate on the Water Valve to be calibrated Calibrating Plain Carbonated Water Modules Fill to 200 ml Graduated Cylinder CALIBRATION MAINTENANCE 4 Set the Timer to the ON po...

Page 13: ...r reference The water flow rate was set to 73 93 ml sec which makes the final syrup flow rate 14 79 ml s The finished drink flow rate dispensed will be 88 72 ml s NOTE A B C D Increase Decrease A Flow...

Page 14: ...ttach outer nozzle by pushing up and twisting clockwise until snug 5 Enter the brand calibration screen from the previous section steps 1 2 6 Set the Timer to the ON position and select Time as the Op...

Page 15: ...The following is a set of instructions on how to set up this feature If no sold out sensor is assigned then the Auto feature acts the same as the Ready feature NOTE Sold Out Features 1 From the Servic...

Page 16: ...d scroll down and press Accept to confirm and save changes It may take up to 10 minutes for a network connection to be established NOTE 2 From the service menu select Network Status to change connecti...

Page 17: ...into open port on the side of the PCB controller box 4 Access the Service menu on the left side screen and tap the Data Management button Update Brands Flavors 1 Create a USB drive with the updated b...

Page 18: ...ment aux lignes directrices de la NSF L exploitant de l quipement doit en assurer l entretien continu comme indiqu dans ce manuel ou selon les lignes directrices du minist re de la Sant afin d assure...

Page 19: ...n Box on page 21 Pull out unit if applicable and clean behind and underneath Check for any loose components or noises Other Supplies Needed 1 Clean cloth towels 2 Bucket 3 Extra nozzle 4 Sanitary glov...

Page 20: ...in the sanitizing solution using the nozzle brush dip the brush in the nozzle sanitizing solution and thoroughly brush the bottom of the inner nozzle body 6 Rinse the brush in warm 90 110 F 32 2 43 3...

Page 21: ...all sides of the Ice Bin and surface of the aluminum casting 5 Place syrup lines with BIB connectors and any adapters into cleaning solution 6 Activate each valve until lines are filled with cleaning...

Page 22: ...coming CO2 pressure is between 70 psi 0 483 MPa and 80 psi 0 552 MPa 2 Open shutoff fully 3 Remove soda flow control from valve and clean out any foreign material to ensure smooth spool movement Insuf...

Page 23: ...to syrup pump too low 5 Stalled or inoperative BIB pump 6 Kinked line 7 CO2 regulator malfunction 1 Replace syrup BIB as required 2 Open shutoff completely 3 Remove valve from mounting block open shu...

Page 24: ...Wiring harness not plugged in 3 PC board defective 4 Malfunctioning power supply 1 Connect dispenser to power source 2 Plug in wiring harness 3 Replace PC board 4 Check voltage to power supply Check f...

Page 25: ...ck BIB connector if still leaking then re place line 2 Replace o ring 3 Replace defective pump BIB pump fails to restart after bag replacement 1 BIB connector not on tightly 2 BIB connector is stopped...

Page 26: ...12 14 16 2 4 6 16 8 10 12 14 1 3 5 7 9 NC 11 13 15 NC 1 3 5 15 7 9 11 13 BLK WHT GRY BLK PUR BLU GRN YEL ORG RED BRN VIO BLK BLK GRN BLK BRN BLK ORG BLK BLK BLK GRY PUR BLU GRN YEL ORG RED BRN WHT BLK...

Page 27: ...R USB DC BLOCK OK POWER CORD CONDENSATION FAN PRIMARY PCB SECONDARY PCB POUR BUTTON POUR BUTTON GPIO ICE DOOR TRIGGER VALVEHARN RIGHT VALVEHARN LEFT SBC USB A TO USB B DIO PCB POWER SUPPLY USB A MALE...

Page 28: ...SATION FAN PRIMARY PCB SECONDARY PCB GRN DC FAN AC OUT AC VOLT TO SUPPLY GRN BLU BLK DIO PCB BLK RED RED BLK BLK RED DENOTES DEFAULT 3 2 SW2 SWITCH 2 POSITION DOES NOT MATTER 4 ON LANCER PN 06 3289 01...

Page 29: ...29...

Page 30: ...A HOSHIZAKI COMPANY 6655 Lancer Boulevard San Antonio TX 78219 lancerworldwide com...

Reviews: