28
SMART Spectro 2 Operator’s Manual 04.16
6. Press or
to highlight
NO
or
YES
.
CALIBRATE WL
CALIBRATE WL???
PLEASE SELECT
NO
7. Press
*ENTER
to begin the wavelength
calibration.
CALIBRATE WL
CALIBRATE WL???
PLEASE SELECT
YES
During the calibration, the Spectro will display progress updates on the top line of the
display. The microprocessor will move the grating in search of the position that gives a
very specifi c maximum light intensity. The microprocessor will then move the grating a
precise predetermined amount from this position. This precise movement will cause the
grating to be positioned at 546 nm every time. Once calibrated the wavelength displayed
during testing is accurate to ±2 nm. When the wavelength calibration is complete the
display will go back to the System Setup Menu.
BT PRINTER
Bluetooth wireless technology allows communication between the spectrophotometer
and the Mobile Bluetooth Printer (Code 5-0066 only). The Bluetooth feature is always on.