Test Procedures
PROCEDURE
Use universal sample holder.
1. Press
ON
button to turn spectrophotometer on.
2. Scroll to and select
PROGRAMMED TESTS
.
3. Scroll to and select
ALL TESTS
(or another sequence containing
45 Hydrazine
) from
TESTING MENU
.
4. Scroll to and select
45 Hydrazine
from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select
SCAN BLANK
.
7.
Remove tube from Spectro. Use the 1 mL pipet (0354) to add 4 mL of *Hydrazine
Reagent A (4841). Cap and mix.
8. Use the 0.15 g spoon (0727) to add one measure of Hydrazine Reagent B Powder
(4842). Cap and shake vigorously for 10 seconds. Wait 2 minutes for maximum color
development. An undissolved portion of Hydrazine Reagent B may remain in bottom of
tube without adversely aff ecting results.
9. At the end of the 2 minute waiting period, mix, insert tube into chamber, close lid and
select
SCAN SAMPLE
. Record result.
10. Press and hold
OFF
button to turn spectrophotometer off or press
EXIT
button to exit
to a previous menu or make another menu selection.
NOTE: For best possible results, a reagent blank should be determined to account for any
contribution to the test result by the reagent system. To determine the reagent blank,
follow the above test procedure to scan a distilled or deionized water blank. Then follow
the above procedure to perform the test on a distilled or deionized water sample. This test
result is the reagent blank. Subtract the reagent blank from all subsequent test results
of unknown samples. It is necessary to determine the reagent blank only when a new lot
number of reagents is obtained.