60
SMART Spectro 2 Operator’s Manual 04.16
5. Press
or
to scroll to the user test to
be edited.
EDIT USER TEST 535NM
101 USER TEST 1
102 USER TEST 2
103 USER TEST 3
104 USER TEST 4
105 USER TEST 5
106 USER TEST 6
107 USER TEST 7
108 USER TEST 8
6. Press
*ENTER
to select the user test to be
edited.
NOTE: The menu allows a test to be
renamed, a wavelength to be selected,
reacted standards to be scanned and the
numerical format of the displayed test
result to be selected. After editing any
one of these menu choices the display
will return to this menu. Any menu choice
can be edited at any time by selecting
it. The normal procedure would be to
start with
NAME THE TEST
, then
SELECT
WAVELENGTH
, then
NEW CALIBRATION
,
and then
FORMAT RESULT
.
102 USER TEST 2 535NM
NAME THE TEST
SELECT WL
NEW CALIBRATION
FORMAT RESULT
NAMING THE TEST
A NAME can be up to 14 characters long. The menu choices for each character are 26
letters A to Z, ten numerals 0 to 9, a space, a dash, a decimal point, and a ! as a terminator.
Selecting the terminator indicates the end of the name and stores the name. The
terminator, !, is the fi rst menu choice since it is the one character that will always have to
be selected. It is before the letter A.
DO NOT CHOOSE THE ! UNTIL THE NAMING OF THE TEST
HAS BEEN COMPLETED
. The terminator should be selected following the end of the name.
It must be the last character selected.