Test Procedures
PROCEDURE
Use universal sample holder.
1. Press
ON
button to turn spectrophotometer on.
2. Scroll to and select
PROGRAMMED TESTS
from menu.
3. Scroll to and select
ALL TESTS
(or another sequence containing
32 Copper DDC
) from
TESTING MENU
.
4. Scroll to and select
32 Copper DDC
from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select
SCAN BLANK
.
7.
Remove tube from Spectro and add 5 drops of *Copper 1 (6446). Cap and mix.
Solution will turn yellow if copper is present.
8. Insert tube into chamber, close lid and select
SCAN SAMPLE
. Record result.
9. Press and hold
OFF
button to turn spectrophotometer off or press
EXIT
button to exit
to a previous menu or make another menu selection.
NOTE: The reaction may stain the tubes. Scrub the tubes thoroughly after each use.
COPPER
, DDC
SMART Spectro 2 Test Procedures 05.02.16