Test Procedures
PROCEDURE
Use universal sample holder.
1. Press
ON
button to turn spectrophotometer on.
2. Scroll to and select
PROGRAMMED TESTS
.
3. Scroll to and select
ALL TESTS
(or another sequence containing
79 Phosphate H
) from
TESTING MENU
.
4. Scroll to and select
79 Phosphate H
from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select
SCAN BLANK
.
7.
Remove tube from Spectro. Use the 1.0 mL pipet (0354) to add 2.0 mL of
*VM Phosphate Reagent (4410). Cap and mix. Wait 5 minutes for full color
development.
8. After 5 minute waiting period, mix, insert tube into chamber, close lid and select
SCAN
SAMPLE
. Record result.
9. Press and hold
OFF
button to turn spectrophotometer off or press
EXIT
button to exit
to a previous menu or make another menu selection.
PHOSPHATE
, HIGH RANGE
SMART Spectro 2 Test Procedures 05.02.16