78
SMART Spectro 2 Operator’s Manual 04.16
BATTERY OPERATION
The SMART Spectro 2 can be run using battery power. The battery pack consists of a
rechargeable Ni-metal hydride battery pack. The Battery Pack (Code 2000-BP) and
Battery Charger (Code 2000-BC) are not included in the standard meter package and
must be purchased as accessories. The power cable and AC adapter also used to charge
the battery pack are included with the SMART Spectro 2.
CHARGING THE BATTERY PACK
1. Connect the Power Cable to the AC Adapter.
2. Connect the AC Adapter to the Battery Charger (2000-BC).
3. Connect the Battery Charger to the Battery Pack (2000-BP).
4. Plug the Power Cord into an outlet. The red light will illuminate. When the battery is
fully charged the green light will illuminate. A full charge will require about 5 hours.
RUNNING THE SMART SPECTRO 2 FROM THE BATTERY PACK
1. Connect the Battery Pack to the SMART Spectro 2.
CAUTION: Do not attempt connect the AC Adapter directly to the Battery Pack. The
connectors will not fi t. Do not force them. The AC Adapter must be connected to the
Battery Charger which is then connected to the Battery Pack.
MAINTENANCE
CLEANING
Clean with a damp, lint-free cloth.
DO NOT ALLOW WATER TO ENTER THE SPECTROPHOTOMETER CHAMBER OR ANY OTHER
PARTS OF THE METER.
LAMP REPLACEMENT AND REALIGNMENT
The quartz halogen bulb in the SMART Spectro 2 has an approximate life of 1000 hours. If
a test is performed and there is no response and the unit is receiving adequate power, the
light bulb may need to be re-secured or replaced. With the meter power ON check the rear
lamp access panel to see if any light acan be observed through the vents. If the lamp is not
ON it is either loose or burnt out.
www.
GlobalScientificSupply
.com
GLOBAL SCIENTIFIC SUPPLY
The Laboratory Supply Company!