73
|
www.lakeshore.com
4.5.2.6 Analog Output
4.5.2.7 Digital I/O
SENSe:CRYO
<Boolean 0 = FALSE, 1
= TRUE>
Enables or disables
cryogenic operation.
SENSe:PCOMpensation?
Returns whether probe
frequency compensa-
tion is enabled.
SENSe:PCOMpensation
<Boolean 0 = FALSE, 1
= TRUE>
Enables or disables
probe frequency com-
pensation.
Command
Parameter(s)
Description
SOURce:AOUT?
Returns the current analog output
configuration designator.
SOURce:AOUT
<XRAW | YRAW | ZRAW
XCORrected |
YCORrected |
ZCORrected |
MCORrected>
Sets the source for the analog output:
XRAW = Raw X
YRAW = RAW Y
ZRAW = Raw Z
XCORrected = Corrected X
YCORrected = Corrected Y
ZCORrected = Corrected Z
MCORrected = Corrected magnitude
Command
Parameter(s)
Description
DIGital:IDATa?
Returns the integer representation
of the input states.
DIGital:INPut#:
FUNCtion?
Returns the present function for
input #.
DIGital:INPut#:
FUNCtion
<NONE | MRESet>
Configures the function of the input
#.
NONE does no action.
MRESet resets the max and min
hold values when the input transi-
tions high.
DIGital:OUTPut#:
FUNCtion?
DIGital:OUTPut#:
FUNCtion
<MANual | FCRamping>
Configures the function of the ouput
relay #.
MANual sets the state of the output
according to the OSETtting value.
FCRamping closes the output relay
when the field control setpoint is
ramping to a new value.
DIGital:OSETting?
Returns the integer representation
of the output setting.
DIGital:OSETting
<integer representa-
tion>
Configures the output setting which
determines the states of outputs
assigned the MANual function.
Command
Parameter(s)
Description