Interface Commands:
DIGital:INPut#:FUNCtion,
DIGital:IDATa?
The input states are sampled every 10 ms in sync with the field reading minimum
averaging window. The sampling occurs during the middle of the reading (5 ms off-
set). Querying the input states always returns the most recent value.
3.5.5.3 Outputs
The
DIGital:OUTput#:FUNCtion
command is used to configure a specific out-
put relay or through the Digital I/O menu option in the Settings menu. The
DIGital:OSETting
command is used to configure the state of an output pin that
is configured for manual mode.
Interface Commands:
DIGital:OUTput#:FUNCtion,
DIGital:OSETting
FIGURE 3-23
Input diagram
FIGURE 3-24
Input sync diagram
GND
INPUT 2
INPUT 1
LOW (0)
-5 V
+35 V
+4 V
+1 V
LOW (0)
1
2
3
4
5
6
HIGH (1)
5 V
–
+
–
+
HIGH (1)