6 OPERATION
92
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
6.9.5.1.
Digital Inputs
Configure Digital Inputs as needed per application. Digital input triggers can be defined
as either a rising edge or a falling edge.
Falling Edge
Rising Edge
The table below lists all available Digital Inputs.
Digital Input Options
Digital Input
Comments
Start / Stop
Not in-cycle and receive input then start. If in-cycle and
received input, then stop cycle.
Sniff / Vac
Default is vacuum.
Internal Cal
Set to internal calibration mode; then, initiate a calibration.
Zero
Toggle Zero function to On / Off.
Inlet Vent
Toggle Vent valve Open / Closed.
Clear
Reset error or warning message.
Int Cal Check
Set to internal calibration mode, then start Cal Check.
Gross
Set test mode to Gross.
Fine
Set test mode to Fine.
Ultra
Set test mode to Ultra.
None
No function.