7 TROUBLESHOOTING
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
121
Fault Level Type
Code Short Description
Long Description
Possible Solution (s)
5 -
Warning
–
Detector
not in
normal
condition
?WA
97
Temperature too high
•
Ambient temperature too
high
•
Fan defective
•
Ensure leak detector is
used in the required
temperature tolerance
•
Replace the fan
?WA
98
Temperature too low
Ambient temperature too low
Ensure leak detector is used
in the required temperature
tolerance
?WA
230
Auto. cal. required
Calibration checking results:
defective detector calibration
Launch an autocalibration
?WA
255
Out start condition
•
Leak detector cannot start
because the use conditions
at startup are not met
•
Ambient temperature too
low
Ensure leak detector is used
in the required temperature
tolerance