4 DESCRIPTION
10
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Description
The TITAN VERSA Leak Detector can detect helium and hydrogen tracer gases. It is
suited for leak testing according to the vacuum and sniffer methods.
•
When leak testing according to the sniffer method, the test object is pressurized
with a tracer gas. The optional sniffer line can be used to sniff the test object and
measure the tracer gas escaping through a leak.
•
There are two different options for leak testing according to the vacuum method:
o
The test object can be evacuated, and its outside sprayed with tracer gas.
This type of leak test is often
referred to as “Outside
-
In” testing.
This procedure has the advantage that leaks can be localized.
o
The test object can be pressurized with tracer gas and tested inside a vacuum
chamber. To accomplish this, the TITAN VERSA is connected to a vacuum
chamber. The device measures the overall leakage rate of the test object.
This type of leak test is often
referred to as “Inside
-
Out” testing.
The device has three sensitivity levels for vacuum method:
“
GROSS,"
“
FINE,
”
and
“
ULTRA." Depending on the pressure in the inlet flange and the user settings, the
TITAN VERSA automatically selects the most sensitive level.
The TITAN VERSA Leak Detector includes the following main components. Check the
scope of delivery of the product for completeness after receipt.
•
Main leak detector assembly
•
Documentation packet
o
TITAN VERSA Operations and Maintenance Manual
o
TITAN VERSA Quick Start Guide
o
Leak standard certificate
o
Quality certificate
o
USB drive with documentation
•
Lifting Packet (not included in VERSA C)
o
TITAN VERSA Lifting Guide
o
Two lifting eyes
o
Lifting sling
•
Maintenance Kit (see Maintenance Tools and Parts for contents)
o
Maintenance tools
o
Spare parts
o
Oil drain and fill accessories (wet pump versions only)