8 MAINTENANCE AND SERVICE
144
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
1.
Disconnect power from leak detector.
2.
Remove covers on module enclosure (see Removing Covers).
3.
Remove valve electrical (DIN) connectors with Philips screwdriver.
4.
Remove four long screws on valve coil assembly.
5.
Remove valve coil and plunger (shown above).
6.
Replace with new valve coil and plunger.
7.
Reconnect items per instruction above.
TITAN VERSA Turbo Pump Parts
Part Number
Description
T10034
Oil Wick SplitFlow 80
Replace the Oil Wick in SplitFlow 50
The turbo pump is filled with operating fluid upon the delivery of the SplitFlow 50.
Replace the oil wick cartridge every 12000 hours or 24 months (whichever comes first).
WARNING:
Danger of poisoning due to contact with harmful
substances
The oil wick cartridge and parts of the pump may be contaminated with toxic
substances that are contained in the pumped media.
•
Dispose of the oil wick cartridge as stipulated by local regulations.
•
Prevent health hazards or environmental pollution from contamination by
means of suitable safety precautions.
•
Decontaminate affected parts prior to executing maintenance work.