2 SAFETY
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
5
DANGER:
While the device was built according to recognized and
advanced safety regulations, improper use can result in danger to life
and limb of the operator or other persons and damage to the device
and other property.
Electrical Dangers
Considerable voltages arise inside the device. Touching parts where electrical voltage
is applied can result in death.
➔
Disconnect the device from the power supply prior to any installation and
maintenance work. Ensure the electric power supply is reconnected with
authorization.
Testing live parts results in danger to life.
➔
Before starting a leak test, disconnect electrically-operated test objects from the
power supply. Ensure the electric power supply is reconnected with
authorization.
The device contains electric components that can be damaged from high electric voltage.
➔
Before connecting the device to the power supply, make sure that the supply
voltage specified on the device is the same as the local power supply.
Dangers from liquids and chemical substances
Liquids and chemical substances can damage the device.
➔
Adhere to restrictions of use (see Technical Data).
➔
Do not suck up liquids with the device.
➔
Never contact toxic, caustic, microbiological, explosive, radioactive, or other
harmful substances with the device.
➔
Only clean the device using mild household detergents.
➔
Hydrogen and air form a highly explosive mixture.
➔
Do not use tracer gases with a hydrogen concentration of >10%.
➔
Only use the device outside potentially explosive areas.
➔
Do not smoke near device. Do not subject the device to open fire and avoid
sparking.
Danger from suction on inlet flange
In vacuum leak test operating mode, there is strong suction on the inlet flange of the
device. If parts of the body or other objects are in the vicinity of the inlet flange, they may
be sucked in.
➔
Always use the provided inlet screen filter.
➔
Always keep parts of the body or other objects away from the inlet flange.