8 MAINTENANCE AND SERVICE
136
TITAN VERSA Operations and Maintenance Manual
LACO Technologies
Filters
Fan filters should be checked regularly and replaced every 3000 hours or 12 months
(whichever comes first). Compact units have one fan filter (shown below) inside the
main enclosure. Tower and horizontal units have an additional pump fan filter located
in the pump sections of the leak detector (shown below).
8.11.2.1.
Replacing Filter
1.
Remove necessary covers to access fan.
2.
Use hand to pop off plastic guard on fan.
3.
Replace filter (center filter to avoid major gaps).
4.
Replace plastic guard.
8.11.2.2.
Filter Locations
Inlet Fan Filter (all models)
Pump Fan Filter
–
Horizontal Configuration