5 INSTALLATION
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
23
Figure 2:
Bronze Filter Installation
The filter should be cleaned with clean, dry air every three months and replaced every
18 months.
5.2.1.2.
Optimized Vacuum Test Performance Recommendations
•
Verify test setup has proper hoses, connections, and valves in accordance with
good vacuum practice.
•
Use pipe with a diameter equal to the diameter of the detector's inlet. The pipes
should be as short as possible and completely sealed.
•
Do not use plastic hoses such as compressed air pipes.
•
Check that the connected part/installation is impermeable to tracer gas.
•
Test only clean, dry parts or installations with no trace of water, vapor, paint,
detergent, or rinsing products.
•
Test that the entire line is completely sealed when the detector is attached to the
pumping circuit to ensure the connections (pump, pipe, valves, etc.) are correct.