Appendix E:
Quick Chart
Model
30441_
30448_
30461_
30468_
Type
C1
C1
C1
C1
Cabinet Size (in feet)
4
4
6
6
Sash Opening (inches)
10
8
10
8
Starting Serial #
1
1403_
1403_
1403_
1403_
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
Airflow Volume (CFM)
400
323
570
463
Concurrent Balance Volume (CFM)
480
387
684
556
Recommended Duct Vacuum (WC)
.4
.4
.4
.4
Supply HEPA Data
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Labconco P/N
3838511
3838511
3838513
3838513
Fluorescent/UV Lamp Data
Fluorescent Lamps (2 each)
F32T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
UV Lamp
G30T8
G30T8
G30T8
G30T8
1.
The primary serial tag is on the lower outside edge of the front dress panel.
The secondary serial tag is located on the front of the electronics module on
the top right side of the cabinet. The first two digits of the serial number are
the year of production; the next two are the month. The next 5 digits are the
sequence of production, and the letter following the serial number is the
revision level of the cabinet.
2.
Each motor must be programmed by Labconco for the appropriate width
cabinet.