Chapter 3: Getting Started
Product Service 1-800-522-7658
14
Original instructions
Exhaust Connection Configuration
Note:
The Axiom is configured at the factory to operate
unconnected
to an
exhaust system. When operating in this mode, the cabinet ignores the inlet
relief valve position.
If you intend to connect the cabinet to an exhaust
system, YOU MUST RECONFIGURE IT FOR THIS INSTALLATION.
Note:
If the Axiom is connected to an exhaust system, and you choose to
disconnect it, to exhaust its HEPA-filtered air back into the laboratory,
YOU
MUST RECONFIGURE IT FOR THIS INSTALLATION. These
instructions follow.
Connecting the Axiom to an Exhaust System
a)
Mechanical configuration
1.
Using an appropriate ladder or platform, remove the exhaust
cover panel(s) on the top right side of the exhaust cover, and
install them over the hole(s) on the left side. Hand tighten the
wing nuts to secure it, as shown in figure 3-2.
2.
Connect the Axiom’s connection collar to the exhaust system,
ensuring the connection meets all appropriate codes and
regulations.
Exhaust cover installed over
exhaust outlet in left side of cover
top
Exhaust cover removed from the
right side of cover top
Figure 3-2