Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
50
Original instructions
Check all service valves, if so equipped, for proper operation.
Check the UV and fluorescent light hourmeters, and record their
readings in an operational log.
All weekly activities.
Semiannually or Annually
Have the cabinet re-certified by a qualified certification technician.
All monthly activities.
Service Operations
Center Work Surface Removal:
Note:
The work surface must be thoroughly decontaminated before
removing it from the cabinet.
1.
Lift either of the wing work surfaces up by grasping the knob handles at
either front corner.
2.
This will lift up the center work surface. Lift the front edge up, pivoting
the work surface on its back edge until the front edge of the work surface
engages the catch located on the back wall of the work area.
3.
Reinstall the work surface by supporting it while lifting up on the catch to
release the work surface. Gently lower the center work surface, being sure
to engage the pins on the front corners of the work surface with the holes
in the grille, as shown in Figure 6-1.