Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
55
Original instructions
Storage
If the biosafety cabinet is to be left unused for more than one month, it should
be prepared for storage.
Note: The cabinet should not be stored in areas of excess humidity or
temperature extremes. If the cabinet is moved during storage, it must be
recertified before use.
1.
Close the sash completely and seal the bottom edge and the exhaust outlet
with plastic sheeting.
2.
Unplug the cabinet.
3.
Ensure that the cabinet will not be moved or disturbed while being stored.