Chapter 3: Getting Started
Product Service 1-800-522-7658
20
Original instructions
3.
The screen will now prompt you for the
password; it is:
Light
UV Light
Timer
Timer
OK
Note: Any other key sequence will return you to the Attention screen.
4.
The first screen allows you to select
whether the Axiom is connected to an
exhaust system or not.
IT IS
CRITICAL THAT THE
THIS
CABINET HAS NO EXHAUST
CONNECTION
OPTION IS
SELECTED
. Press
OK
to continue.
5.
The third screen allows you to set the sash
height at either 8 or 10 inches.
IT IS
CRITICAL THAT THE YOU LEAVE
THE SASH HEIGHT AS IT WAS SET
AT THE FACTORY, UNLESS YOU
WISH TO RECERTIFY THE CABINET
AT ITS NEW SETTING
.
6.
The fourth screen allows you to select
whether the Axiom has a UV light or not.
IT
IS CRITICAL THAT THE YOU LEAVE
THE UV LIGHT CONFIGURATION AS
IT WAS SET AT THE FACTORY
. Press
OK
to return to the first Configuration
screen.
The unit is now properly configured for operation connected to an exhaust
system. In this mode, the unit will display an exhaust alarm if the inlet relief
valve closes during operation.
NOTE: If you ever want to disconnect the Axiom from the exhaust system,
and use it in a recirculating mode, you must reconfigure it as such.