Chapter 5: Using The Cabinet
Product Service 1-800-522-7658
46
Original instructions
Allow the cabinet to operate until the display screen is shown.
Warm up screen-
Connected to an exhaust system
no exhaust system
Display screen
Wash hands and arms thoroughly with germicidal soap.
Wear appropriate personnel protective equipment (PPE).
Wipe-Down
Raise the sash to its full open position (approximately 22.6 inches or
574 mm). Mute the alarm by depressing the "OK/Mute" switch.
Wipe down the interior surfaces of the cabinet with 70% ethanol, or a
suitable disinfectant, and allow to dry.
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload
the cabinet.
Do not obstruct the front, side, or rear return air grilles.
Large objects should not be placed close together.
Slowly close the sash until it is in the correct operating position.
After loading the cabinet, wait two to three minutes to purge airborne
contaminants from the work area.