Chapter 5: Using The Cabinet
Product Service 1-800-522-7658
35
Original instructions
Alarm Screens
Sash is too high
The sash is open too far for safe operation.
Airflow Alerts
The airflow patterns in the cabinet have changed, resulting
in a sudden change in either motor speed. This is most
likely due to a blockage of the grille or the exhaust filter
outlet. It may also be caused by removal of the work
surface while the cabinet is in operation.
Exhaust Fan Alarm
There has been a failure of the exhaust fan, or its
communication to the cabinet’s control board. DO NOT
USE THE CABINET UNTIL THE PROBLEM HAS
BEEN CORRECTED.
Exhaust System Alarm-Active Protection Mode
When connected to an exhaust system, and the flow drops
below acceptable limits, this screen will appear. When the
timer reaches 0, the supply blower will shut off, and the
exhaust fan will continue to operate for 10 more seconds,
protecting the operator, until the countdown reaches 0.
Exhaust System Alarm
After the countdown reaches 0, the exhaust fan will shut
off and this screen will be displayed. If this alarm is
displayed, the Axiom blower will need to be turned off,
and then back on again to reset the alarm. Frequent
Exhaust System Alarms indicate unstable or insufficient
exhaust flow, and must be corrected.
System Error
The motor and display circuit board are not communicating
properly. DO NOT USE THE CABINET UNTIL THE
PROBLEM HAS BEEN CORRECTED.
Figure 5-4b
Figure 5-4f
Figure 5-4a
Figure 5-4c
Figure 5-4d
Figure 5-4e