Chapter 3: Getting Started
Product Service 1-800-522-7658
17
Original instructions
7.
The fourth screen allows you to select whether the
Axiom has a UV light or not.
IT IS CRITICAL
THAT THE YOU LEAVE THE UV LIGHT
CONFIGURATION AS IT WAS SET AT THE
FACTORY
. Press
OK
to return to the first
Configuration screen.
The unit is now properly configured for operation connected to an exhaust
system. In this mode, the unit will display an exhaust alarm if the inlet relief
valve closes during operation.
NOTE: If you ever want to disconnect the Axiom from the exhaust system,
and use it in a recirculating mode, you must reconfigure it as such.