The following are instructions for mounting a cross support:
1. Level the base cabinets and the work surface. The filtered hood should be placed
flush with the back of the work surface as shown in Figure 4-2.
2. Scribe a line on the wall or back of the base cabinet to locate the support under
the work surface.
3. Mount the support by attaching it to the wall or base cabinet.
4. Place the filtered hood on top of the work surface and cross support.
Step 2
Prior to connecting any electrical wiring to the filtered hood structure, refer to the filtered
hood serial tag for the proper electrical requirements of your specific model. Serial tag
location is shown in Figure A-1 of
Warning
:
The building electrical supply system for Protector Hoods should
include overload protection. A switch or circuit breaker should be in close
proximity to the equipment and within easy reach of the operator. The switch or
circuit breaker is to be marked as the disconnecting device for the equipment.
Consult the current version of NFPA 70
®
, NEC
®
for proper installation.
Avertissement
: Le système d’alimentation électrique de la Hotte Protecteur doit
inclure la protection contre la surcharge. Un commutateur ou disjoncteur doit
être tout près de l'équipement et à portée facile de l'opérateur. Le commutateur
ou le disjoncteur doit être marqué comme l'appareil débranchant pour
l'équipement. Consultez la version actuelle de la norme NFPA 70
®
, NEC
®
pour une
installation correcte.
The serial tag, model number, serial number, and electrical connection boxes are
accessible from the front of the filtered hood by removing the front panel. Check the
serial tag behind the front panel for voltage verification. The number of circuits varies
depending on the model. All of the electrical connections for the individual duplexes are
terminated at the single point internal junction box for hook-up by a qualified electrician.
If needed, the individual duplexes can be converted for instant attachment to a wall
outlet by a qualified electrician by ordering 115V harness 9582710 or 230V harness
9582720. The main power to the control panel, fans, and lights uses less than 3 amps
and a main power cord is included for instant attachment to a wall outlet. The single
point internal junction box is used for the connection of the duplex outlets. Refer to the
wiring diagram for the filtered hood in
The filtered hood is required to be grounded to the MAINS protective earthing ground
for safe operation. Using a ring terminal sized for a #10-24 machine screw, connect the
MAINS ground conductor to the grounding lug marked with the protective earthing
symbol. Only MAINS ground conductors should be connected to the protective earthing
ground lug, no other conductors should be connected to this grounding lug. Using wire
nuts, connect the MAINS supply conductors to the filtered hood supply wires. Ensure
that the wires are connected as per the appropriate wire color codes for the input
voltage. For 115V Phase (Hot) is black and Neutral is white, for 230V Phase1 is brown
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...