Step 5
Start up the filtered hood. If the filtered hood is in alarm, check eGuard for the
explanation. If eGuard says the base frame has been replaced, press the
[mute]
button
on the command module. If there is an alarm other than the replacement message,
disconnect the power, re-check all electrical connections and try again.
When properly working, reinstall the torx screws and screw caps and resume using the
filtered hood.
Step 6 (if needed)
Recalibrate the filtered hood by the following the steps in