104
The Protector Echo Floor-Mounted Filtered Hood models weigh between 700 to 1500
lbs. (318-675 kg). The multiple shipping pallets allow for lifting with a mechanical lift
truck or floor jack. If you must lift the filtered hood manually, follow safe-lifting
guidelines.
Les modèles filtrée protège-capot Echo monté au plancher pèsent entre 700 à 1500
livres. (318 à 675 kg). Les multiples palettes d'expédition permettent de levage avec un
chariot élévateur mécanique ou prise de parole. Si vous devez soulever la hotte
manuellement, suivre les directives safe-levage.
Unpacking
Carefully remove the shrink-wrap and/or outer carton and inspect the product for
damage that may have occurred in transit. If the product is damaged, take pictures of
the product and the outer packaging, and notify the delivery carrier immediately. Retain
the entire shipment, including outer packaging, intact for inspection by the carrier.
Note:
Do not return goods without the prior authorization of Labconco. Unauthorized
returns will not be accepted. If your hood was damaged in transit, you must file a claim
directly with the freight carrier. Labconco Corporation and its dealers are not
responsible for shipping damages.
Do not move the filtered hood by tilting it onto a hand truck.
Ne pas déplacer le capot filtrée en l'inclinant sur un camion de main.
Remove from Shipping Skid
Note:
Leave the filtered hood attached to its shipping skid until it is as close to its final
location as possible. Move the hood by using a suitable floor jack, or by placing a
furniture dolly underneath the skid. Do not move the hood by tilting it onto a hand truck.
After you verify the filtered hood components, move the hood to the installation location.
Follow the steps listed to remove the shipping skids from the lower base and upper
cabinet.
1. Remove the side panels by unscrewing the Phillips screws.
2. Find the hardware (bolts, washers, nuts) that attach the filtered hood to the skid
and remove the hardware.
Lower Base Installation
Remove the left and right side panels for accurate placement of the lower cabinet
assembly. Remove both vertical-rising sashes shipped on the same skid as the lower
base. Place the sashes in a location where they will not get lost or damaged. Position
the lower base on a level floor surface or an optional purchased floor surface. If an
optional floor surface is purchased, it should be positioned on the existing floor prior to
placement of the lower cabinet installation.
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...