69
9: Accessories
This section details the available field-installable accessories and approved
modifications for the filtered hood. These include the addition of replacement
Neutrodine carbon filters, HEPA filters, work surfaces, base stands, base stand shelves,
seismic supports, storage cabinets, service fixtures, electrical duplex outlets, sash
stops, distillation grids and sash reduction wing kits.
Filters
The filtered hood model sizes, filter compartments, and number of filters are listed
below for installation convenience.
Hood
Size
Filter
Columns
Required No.
of Neutrodine
Filters
Total
Neutrodine
Filter Weight
Optional No.
of HEPA
Filters
Total HEPA
Filter
Weight
3 FT
1
2
50 lbs.
1
10 lbs.
4 FT
2
4
100 lbs.
2
20 lbs.
5 FT
3
6
150 lbs.
3
30 lbs.
6 FT
4
8
200 lbs.
4
40 lbs.
8 FT
5
10
250 lbs.
5
50 lbs.
Work Surfaces
A required work surface is available to support the filtered hood. The black epoxy work
surfaces are contoured to fit the dimensions of the filtered hood to contain spills. May be
ordered with a pre-cut 6-inch x 3-inch oval cupsink cutout.
Catalog #
Description
9577420
Neutrodine molecular carbon filter (universal)
9577421
Optional HEPA filter
9577422
Pre-filter, located in Base Frame, Standard
9577423
Pre-filter, located in Base Frame, Optional: Inorganic Acids
9577424
Pre-filter, located in Base Frame, Optional: Ammonia
Table 9-2
Table 9-1
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...