27
and Phase2 is blue. Refer to the wiring diagram for the filtered Hood in
It is recommended all wiring for the filtered hood be performed by a licensed
electrician and conform to all local codes.
Il est recommandé que tout le câblage de la hotte filtrée soit effectué par un
électricien agréé et conforme à tous les codes locaux.
Step 3
The filtered hoods with service fixtures have been plumbed from the valve to the hose
connector or gooseneck for your installation convenience. Supply tubing shall be
provided by the qualified installer. Tubing can enter the filtered hood from above,
through the back, or through the work surface to make these connections to the service
fixtures.
Note:
Inspect all fittings for leaks. Tighten the fittings slightly if necessary.
Remarque:
Inspectez tous les raccords pour les fuites. Serrez légèrement les raccords
si nécessaire.
Caution:
Do not use oxygen with any standard service fixture. Contact Labconco
Customer Service for oxygen fixture information.
Prudence :
Ne pas utiliser de l'oxygène avec l'accessoire de service standard.
Contacter le Service Clientèle de Labconco pour les informations d'accessoire
d'oxygène.
Should access to the filtered hood plumbing fixture bodies be required, remove the
service access plate on the filtered hood front corner posts by loosening their individual
screws (see item 8 of Table A-1 in
). The valve body will now be
fully exposed for any service work that may be necessary. The service fixtures supplied
on the filtered hood are designed for use with the following services:
Air
Hot Water
Vacuum
Cold Water
Natural Gas
– See
cautions below
Warning:
Contact Labconco Customer Service directly before using any service other
than those listed above in these valves to assure full compatibility.
Avertissement:
Contacter le Service Clientèle de Labconco directement avant d'utiliser
n'importe quel service autre que ceux énumérés au-dessus dans ces soupapes pour
assurer une pleine compatibilité.
Caution:
Natural gas should be used only in the service fixture that has been pre-
plumbed with brass tubing. Sulfur content of the gas could cause deterioration of
standard copper supply lines.
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...