100
Update:
Connecting to your filtered hood with eGuard
Another page with options for manipulation is the update section of eGuard. Your
phone, tablet or laptop still needs to be connected to the filtered hood:
1. Open the network selector on your device.
2. Select the Wi-Fi network:
GFH-XXXXX-XXXX
(the
XXXX
is the serial number of the filtered hood).
3.
Password is:
GFH-2018
4.
Open an internet browser (Google Chrome, Mozilla Firefox… etc.) and
enter
the following address in the browser menu
Where a username and password is required:
a. Username:
ach
b. Password:
ach
For connecting the filtered hood to a laptop with Ethernet:
1. Connect an Ethernet cable to the laptop from the command module.
2. Configure the Ethernet parameter of computer:
a. Protocol TCP/IP V4
b. IP address: 192.168.0.205
c. Subnet mask: 255.255.255.0
3. See Figure 6-3 for more details.
4.
Open an internet browser (Google Chrome, Mozilla Firefox… etc.) and
enter the
following address in the address menu:
Where a username and password is required:
a. Username:
ach
b. Password:
ach
This will open a webpage that is the eGuard
update
home screen. The homepage has
three sections called INFO, LOGS and SETTINGS. INFO is the only section that makes
).
INFO
More filtered hood adjustments are available in the INFO (update) section. Adjustments
like fan and sensor settings, factory reset, server configuration, Ethernet and
maintenance settings.
Fan Settings
In this section the rotation fan speed set points can be
managed.
Working Setpoint
is the rotation speed of the
fans when the sash is open. When the sash is open, the
fans will be rotating at this speed. There is also
Closed
Setpoint
which is the rotation speed of the fans when
the sash is closed. When the ventilation is on but the
sash is not open, the fans will rotate at this set point.
Force Start
authorizes the start of
Figure D-13
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...