59
Fan Replacement
Fans will need replacement when damaged or if failure occurs in the fan’s motor or
solvent sensor.
This section will cover how to:
Remove front panel
Identify your model and order new fans
Unpack new fan boxes
Access the old fan boxes
Remove and dispose of old fan boxes
Insert the new fan boxes
Replace secondary filters in filtration column
Reinstall front panel and start up
The following tools are required to replace the equipment:
6 FT ladder
Tape
Marker
The following safety precautions must be followed by all personnel handling the
equipment and filtration columns:
It is recommended you follow all of your existing lab safety protocols as they
apply to this procedure.
2 persons required for this procedure.
Read instructions carefully before attempting.
Do NOT have fans running during this process.
Step 1
Turn the filtered hood off and disconnect the power plug. Remove the front panel (see
Section 8: Front Panel Removal
) of the filtered hood and place nearby but out of the
way of the filtered hood. This will expose the filtration column technology you will be
working with.
Step 2
Identify the filtered hood you have and order the correct quantity for replacing.
Determine your model number which is a M1 (3 FT), M2 (4 FT), M3 (5 FT), M4 (6 FT),
or M5 (8 FT). These numbers are based on how many stacks of technology that you
have on top of your filtered hood from left to right and the filtered hood width.
Look at your filtered hood and count the stacks of technology from left to right
You will have 1, 2, 3, 4, or 5 stacks for 3 FT, 4 FT, 5 FT, 6 FT, and 8 FT filtered
hoods.
Match that number to M1, M2, M3, M4, or M5.
This information will help you follow this procedure when it applies to your model
M1, M2, M3, M4, M5.
Summary of Contents for Protector Echo
Page 78: ...78 Fans and Lights not working...
Page 79: ...79 Vertical sash no longer operates smoothly...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working...
Page 81: ...81 Contaminates outside the filtered hood...
Page 82: ...82 Lights operate but fans will not...
Page 83: ...83 Smart Command Alarms...
Page 84: ...84 If needed contact Labconco to troubleshoot further...
Page 94: ...94 Wiring Diagram Figure C 1...