46
8.
TEDS OPTION
8-1
OUTLINE
By connecting the TEDS (Transducer Electronic Data Sheet) built-in sensors to the WGA-670B-7, the most
appropriate sensitivity registration is automatically available.
Connecting up to 4 sensors of the same standard is available.
8-2
CONNECTION
Connect TEDS signal sensors to the following terminal boards Nos. 21 and 22 on the rear panel as shown in the
following figure.
Take care of the polarity of the TEDS signals and correctly connect the signals. Or, error occurs when reading
data. (Error display ‘E-03’). However, no sensors are damaged by connecting to wrong polarity.
Figure 8-1 Connecting TEDS signals
TEDS(+) Yellow lead wire
TEDS(COM) Blue lead wire