17
Function
Selecting Mode Function Name
Key Operation & WGA-670B
Movement in Function Setting Mode
Functional
Descriptio
n
F-12
Initialize
•Key code enable setting
‘00000’ is indicated.
Here, set the indication to read ‘00628’ or ‘11111’
and press the SET key. The following set values
return to the factory default state. Then, the key
code is enabled. Therefore, when returning the set
value to the factory default the next time, the
above operation is required.
To disable the key code, conduct the function
‘F-12.’
•Key code disable setting
‘00000’ is indicated. Press the SET key as it is.
Then, the following set value returns to the
factory default state.
The key code is disabled. The set value returns to
the factory default stage even if no key code is
entered.
●
Factory default state
Hold mode
0-0-0
Delay time
0 second
Specified time
0 second
Analog output
0 V with 0000 display
10 V with 4000 display
Smoothing 0-0-0
Additional value
0
Auto ZERO Compensation range
0
Determination time
0.0 second
ZERO near ZERO
0
Hysteresis 0
BCD logic
0-0-0
HIGH/LOW limit set value
HIGH limit
+19999
LOW limit
-19999
●
Calibration value and Bridge Exciter value
(BV) are not initialized.
●
The indication ‘00628’ or ‘11111’ is a key code
to prevent calibration value from carelessly
changing due to malfunction.
• A
function
that
returns
the set
values to
factory
default
state.