37
7-4
DETAILS of VARIOUS CONTROL COMMANDS
This section describes details of various control commands and parameters that are following the commands.
(<CR>,<CR><LF> represent termination code.)
(1) [RES]
Contents: Initialize
(Relevant to panel operation F-12.)
Format: RES0628<CR>
Key code (Input ‘0628’ regardless of enable/disable
of the key code.
Parameter: None
Return value format:
ok<CR><LF>
Description & others:
Results of this command are disabled with the power OFF.
(2) [RS-]
Contents:
RS line test
Format: RS-<CR>
Parameter: None
Return value format:
RS-232C<CR><LF>(When normally received)
(3) [SYS]
Contents:
Outputs Model name, etc.
Format: SYS<CR>
Parameter: None
Return value format:
WGA-670B-a,xx.xx,yyyyyyyyy<CR><LF>
Production
No.
Program version
Model
Description & others:
Details of the return value ‘a’ is as follows.
0:
No
option
1:
BCD
option
7:
RS-232C+TEDS
option
Summary of Contents for WGA-670B
Page 3: ......
Page 56: ...51 9 2 OUTSIDE DRAWING 1 With No Option WGA 670B 0 ...
Page 57: ...52 2 With BCD Option WGA 670B 1 ...